Español

ABM-4082 Benchtop Digital Multimeter

Absolute error of my measurement

"Absolute measurement error" service allows you to estimate the measurement accuracy considering uncertainty (error) in the absolute value. It’s important if you take measurements at the range edge when the error value contributes to the value displayed on the device screen.

To get the value range it’d be enough to enter the estimate of the value under measurement into Value input field and to see the range of possible values taking all of the standardized errors into consideration.

Note: a point is used as a separator of the integer and fractional part.

Software delivery

  • Test leads (pair)

  • USB-Host cable.- 1

  • Power Cord

AKTAKOM ABM-4082 Benchtop Digital Multimeter
Click an image to enlarge
Brand: AKTAKOM
User's Guide: Download
RS-232 (Recommended Standard 232) True RMS

Digital Multimeter AKTAKOM ABM-4082 is a benchtop instrument designed for AC and DC voltage and current, resistance and frequency measurements. Its 12 different measurement functions: DCV/ACV, DCI/ACI, Ω2W/Ω4W, frequency/period, diode test, continuity, dB/dBm cover all basic measurement needs.

Retail price : $689.00

Free shipping for orders over $450
on items weighting under 20 lbs
(certain restrictions apply!)

See shipping page for international rates
or contact sales@tmatlantic.com

Almost sold out, 1 left

Buy Add to cart

Notify when back in stock

Digital Multimeter AKTAKOM ABM-4082 is a benchtop instrument designed for AC and DC voltage and current, resistance and frequency measurements. Its 12 different measurement functions: DCV/ACV, DCI/ACI, Ω2W/Ω4W, frequency/period, diode test, continuity, dB/dBm cover all basic measurement needs. GPIB and RS-232 Interfaces provide easy system communication.

Features:

  • 12 different measurement functions: DCV/ACV, DCI/ACI, Ω2W/Ω4W, frequency/period, diode, continuity, dB/ dBm, etc
  • High-brightness vacuum fluorescent display (VFD), 5 1/2 digits 
  • True RMS AC voltage and current measurements, frequency up to 100kHz
  • Maximum reading of 120000 digits
  • DCV measurement accuracy up to 0.01%, resolution of 1μV
  • MAX Measurement rate: 45 times/second
  • Equal accuracy frequency measurement more than 1MHz, minimum resolution of 10μHz
  • Zeroing (REL) mode to eliminate residual reading
  • 2W and 4W resistance measurement modes selectable
  • Build-in mathematics calculation functions: mX+b, %, dB, dBm
  • 512 readings storage and MAX/MIN/AVER/STD statistics
  • HI/IN/LO comparator function
  • GPIB and RS-232 interfaces for convenient system communication
  • Calibration performed without opening the cover
  • 10 sets of measurement states saved and recalled up

DC Voltage (DCV)1000 V
DCV Accuracy±(0.01%+4)
DCV Resolution1 μV
AC Voltage (ACV)750 V
ACV Accuracy±(0.1%+100)
ACV Resolution1 μV
DC Current (DCA)10 A
DCA Accuracy±(0.05%+4)
DCA Resolution0.1 μA
AC Current (ACA)10 A
ACA Accuracy±(0.25%+80)
ACA Resolution0.1 μA
Frequency5 Hz...1 MHz
Resistance100 Ω...100 MΩ
Diode test+
Continuity check+
True RMS+
Min/Max measurement+
Relative measurement+
Special featuresTwo and four-wire resistance measurement
InterfaceRS-232, GPIB
DisplayVFD, 5 1/2 digits (120000 counts)
Power supply100...240 VAC
Dimensions8.8 x 4 x 14 in / 225 x 100 x 355 mm

Specifications

Full-scal reading & reading rate (readings/second)
SpeedSlowMediumFast
Full-scale reading (digits) 1199,999119,99911,999
Reading rate (readings/second)DCV, DCI 3.51345
ACV, ACI 23.822
Ω-2W3.51345
Ω-4W2.71032.9
DCV
RangeMAX.readingResolutionAccuracy(1 year)Input impedance
100mV119.9991μV±(0.02%+8)>10GΩ
1V1.1999910μV±(0.01%+4)>10GΩ
10V11.9999100μV±(0.01%+4)>10GΩ
100V119.9991mV±(0.01%+4)10MΩ
1000V1010.0010mV±(0.01%+4)10MΩ
DCI
RangeMAX.readingResolutionAccuracy (1 year)Load voltage/shunt resistance
10mA11.99990.1μA±(0.05%+8)<0.15V/10Ω
100mA119.9991μA±(0.05%+4)<1.5V/10Ω
1A1.1999910μA±(0.10%+4)<0.3V/0.1Ω
10A11.9999100μA±(0.15%+4)<0.15V/10mΩ
AC voltage (ACV true-rms measurement)
Range100mV1V10V100V750V
MAX.reading119.9991.1999911.9999119.999757.5
Resolution1μV10μV100μV1mV10mV
Accuracy(1 year)10~20Hz±(1.5%+100)
20~50Hz±(0.5%+100)
50~20kHz±(0.1%+100)
20k~50kHz±(0.3%+150)0.3%+150)
50k~100kHz±(1%+150)±(1%+100)
AC current (ACI true-rms measurement)
Range10mA1A10A
MAX. reading11.99991.1999911.9999
Resolution0.1μA10μA100μA
Accuracy (1 year)10~20Hz±(1%+80)
20~50Hz±(0.5%+80)
50~2kHz±(0.25%+80)
2k~10kHz±(2%+80)------
Load voltage/shunt resistance<0.15V/10Ω<0.3V/0.1Ω<0.15V/10mΩ
Resistance (Ω2W/Ω4W)
RangeMAX.readingResolutionMeasurement current Accuracy (1 year)
100.000Ω119.9991mΩ1mA±(0.05%+8)
1.00000kΩ1.1999910mΩ1mA±(0.03%+4)
10.0000kΩ11.9999100mΩ100μA±(0.03%+4)
100.000kΩ119.99910μA±(0.03%+4)
1.00000MΩ1.1999910Ω10μA±(0.03%+4)
10.0000MΩ11.9999100Ω7.0xRx/(10M+Rx)±(0.10%+4)
100.000MΩ119.9991kΩ7.0xRx/(10M+Rx)±(0.50%+8)
Frequency
RangeMAX.readingResolutionAccuracy(1 year)Sensitivity(sine wave)
5~10Hz9.9999910μHz±(0.050%+100)200mV rms
10~100Hz99.9999100μHz±(0.010%+10)40mV rms
100~100kHz99.99990.1Hz±(0.005%+2)40mV rms
100k~1MHz999.9991Hz±(0.005%+2)100mV rms

  • Operating temperature & humidity: 0°C...40°C, ≤90%RH
  • Power supply:
    Voltage: 99V...121V, 198V...242V
    Frequency: 47.5Hz...63Hz
  • Power consumption: ≤ 20 VA
  • Dimensions (W×H×D): 8.8x4x14in/225x100x355mm
  • Weight: 5.5lb/2.5kg
  • Test leads (pair)

  • USB-Host cable.- 1

  • Power Cord

El Multímetro AKTAKOM ABM-4082 es un instrumento de sobremesa diseñado para voltaje y corriente AC y DC, resistencia y mediciones de frecuencia. Sus 12 funciones de mediciones diferentes: DCV / ACV, DNI / ACI, Ω2W/Ω4W, frecuencia / período, prueba de diodos, de continuidad, dB / dBm cubren todas las necesidades básicas de medición. Interfaces GPIB y RS-232 proporcionan la comunicación fácil del sistema.

Características:

  • 12 funciones de medida diferentes: DCV / ACV, DNI / ACI, Ω2W/Ω4W, frecuencia / período, el diodo, continuidad, dB / dBm, etc
  • Alto brillo de la pantalla True RMS AC voltage and current measurements, frequency fluorescente de vacío (VFD), 5 1/2 dígitos
  • True RMS voltaje de corriente alterna y corriente, la frecuencia de hasta 100 kHz
  • Lectura máxima de 120000 dígitos
  • Precisión de la medición DCV de hasta 0,01%, la resolución de 1μV
  • MAX Velocidad de medición: 45 veces por segundo
  • Exactitud igualada de medicion de la frecuencia más de 1 MHz, resolución mínima de 10μHz
  • Modo de Reducción a cero (REL) para eliminar la lectura residual
  • Modos de Selecion a 2W y 4W en la medicion de la resistencia
  • Funciones matemáticas de cálculo: mx + b,%, dB, dBm
  • 512 memorias de lectura y estadisticas de MAX/MIN/AVER/STD
  • Funciones de comparacion HI/IN/LO
  • Interfaces de comunicacion de sistema GPIB and RS-232
  • Calibracion sin tener que abrir la tapa
  • Guardado de 10 conjuntos de mediciones y retorno de memoria

Back to the section
 

Please give us your feedback to help us improve our website

* - required fields

AKTAKOM - the Best in Test Award Winner!
Site map|Privacy statement|Terms of Use & Store Policies|How to Buy|Shipping|Payment|© T&M Atlantic, Inc., 2010-2016