Español

AMM-1037 Digital Multimeter

AKTAKOM AMM-1037 Digital Multimeter
Click an image to enlarge
Brand:AKTAKOM
Warranty:1 year
True RMS

High performance digital multimeter

Retail price :$54.34

Depending on a destination shipping cost may vary.
Free shipping for orders over $450
on items weighting under 20 lbs
(certain restrictions apply!)

See shipping page for international rates
or contact sales@tmatlantic.com

In stock now

BuyAdd to cart

High performance Aktakom AMM-1037 digital multimeter is capable of measuring DCV, ACV, DCA, ACA, resistance, capacitance, frequency as well as diode and continuity test.

AMM-1037 uses dual-integral A/D converter as key point.

Features:

  • LCD display with a backlight, 19999(41/2), auto polarity indication
  • Measuring method: dual slope A/D transfer
  • Sampling rate: approx. 3 times/sec
  • Over range indication: OL or -OL should be displayed
  • Low battery indication
  • Operation environment: 0~40℃,
  • Relative humidity: <80%
  • Power: one 9V battery
  • Dimensions: 7.5 x 3.5 x 1.1 in / 190 x 88.5 x 27.5 mm
  • Weight: approx. 14.9 oz / 422g (including battery)

DC Voltage (DCV)1000 V
DCV Accuracy(0.1%+5)
DCV Resolution0,01 mV
AC Voltage (ACV)750 V
ACV Accuracy(0.8%+25)
ACV Resolution0,1 mV
DC Current (DCA)20 A
DCA Accuracy(0.5%+4)
DCA Resolution0,01 A
AC Current (ACA)20 A
ACA Accuracy(1.5%+15)
ACA Resolution0.1 mA
Frequency2 MHz
Resistance200 MΩ
Capacitance2000 F
Diode test+
Continuity check+
True RMS+
Autorange+
Data hold+
Special featuresConductance measurement, hFE measurement
DisplayLCD, 19999 counts
Power supply9 V battery
Dimensions7.5 x 3.5 x 1.1 in / 190 x 88.5 x 27.5 mm

Accuracy tables

DC Voltage

RangeResolutionAccuracyAbsolute errorValue-errorValue Learn an absolute error of your measurementValue+error
200 mV 0.01 mV (0.1%+5 digits) 0.24999 mV mV
2 V 0.0001 V (0.1%+5 digits) 0.0024999 V V
20 V 0.001 V (0.1%+5 digits) 0.024999 V V
200 V 0.01 V (0.1%+5 digits) 0.24999 V V
1000 V 0.1 V (0.2%+5 digits) 2.4998 V V

AC Voltage

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 mV 0.1 mV (1%+25 digits) 4.499 mV mV
2 V 0.001 V (0.8%+25 digits) 0.040992 V V
20 V 0.01 V (0.8%+25 digits) 0.40992 V V
200 V 0.1 V (0.8%+25 digits) 4.0992 V V
750 V 1 V (1.2%+25 digits) 33.988 V V

DC Current

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 A 0.01 A (0.5%+4 digits) 1.03995 A A
2 mA 0.0001 mA (0.5%+4 digits) 0.0103995 mA mA
20 mA 0.001 mA (0.5%+4 digits) 0.103995 mA mA
200 mA 0.01 mA (0.8%+6 digits) 1.65992 mA mA
20 A 0.001 A (2%+15 digits) 0.41498 A A

AC Current

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 mA 0.1 mA (1.5%+15 digits) 4.4985 mA mA
20 A 10 mA (2.5%+35 digits) 0.84975 A A

Frequency

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
20 Hz 0.001 Hz (0.1%+3 digits) 0.022999 Hz Hz
200 Hz 0.01 Hz (0.1%+3 digits) 0.22999 Hz Hz
2 kHz 0.1 Hz (0.1%+3 digits) 0.0022999 kHz kHz
20 kHz 1 Hz (0.1%+3 digits) 0.022999 kHz kHz
200 kHz 10 Hz (0.1%+3 digits) 0.22999 kHz kHz
2 MHz 100 Hz (0.1%+3 digits) 0.0022999 MHz MHz

Resistance

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 Ω 0.01 Ω (0.4%+10 digits) 0.89996 Ω
2 kΩ 0.1 Ω (0.4%+5 digits) 0.0084996 kΩ kΩ
20 kΩ 1 Ω (0.4%+5 digits) 0.084996 kΩ kΩ
200 kΩ 10 Ω (0.4%+5 digits) 0.84996 kΩ kΩ
2 MΩ 100 Ω (0.4%+5 digits) 0.0084996 MΩ MΩ
20 MΩ 1 kΩ (1.2%+25 digits) 0.264988 MΩ MΩ
200 MΩ 10 kΩ (5%+45 digits) 10.4495 MΩ MΩ

Capacitance

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
20 nF 1 pF (4%+50 digits) 0.84996 nF nF
200 nF 10 pF (4%+50 digits) 8.4996 nF nF
2 F 100 pF (4%+50 digits) 0.084996 F F
20 F 1 nF (4%+50 digits) 0.84996 F F
200 F 10 nF (4%+50 digits) 8.4996 F F
2000 F 100 nF (4%+50 digits) 84.996 F F

Technical specifications:

DCV

Range Accuracy Resolution
200mV (0.1%+5) 0.01mV
2V 0.0001V
20V 0.001V
200V 0.01V
1000V (0.2%+5) 0.1V

Input impedance:
200mV range: 1MΩ
other ranges: 10MΩ;

Overload protection:
200mV range: DC 250V or AC peak value.
Other ranges: DC 1000V or AC peak value

ACV

Range Accuracy Resolution
200mV (1.0%+25) 0.1mV
2V (0.8%+25) 1mV
20V 10mV
200V 100mV
750V (1.2%+25) 1V

Input impedance:
200mV range: 1MΩ
other ranges:10MΩ;

The frequency response of sine wave and triangular wave is 40Hz-1kHz; the frequency response of other waveform is 40Hz-200Hz.

Over-range protection:
200mV range: DC 250V or AC peak value.
Other ranges: DC 1000V or AC peak value

Displaying: True RMS (true value response)
Its normal that some range does not turn to 0 and it does not affect the measuring accuracy.

DCA

Range Accuracy Resolution
200uA (0.5%+4) 0.01uA
2mA 0.0001mA
20mA 0.001mA
200mA (0.8%+6) 0.01mA
20A (2.0%+15) 0.001A

Max. measuring volt drop: 200mV
Max. input current: 20A (within 10 seconds)

Over-range protection:
0.2A/250V fuse on range less than 200mA
20A/250V fast melt fuse on other range.

NOTE:
20A range: ≤5A , continuous measurement is allowed.
5A~20A: continuous measurement time should be less than 10 sec. Time interval should be more than 5 min.

ACA

Range Accuracy Resolution
200mA (1.5%+15) 0.1mA
20A (2.5%+35) 10mA

Max. measuring volt drop: 200mV
Max. input current: 20A (within 10 seconds)

Over-range protection:
≤200mA: 0.2A/250V
20A/250V fast melt fuse

Frequency response: (40~200)Hz

Displaying: True RMS

20A range: ≤5A , continuous measurement is allowed.
5A~20A: continuous measurement time should be less than 10 sec. Time interval should be more than 5 min.

Resistance (Ω)

Range Accuracy Resolution
200Ω (0.4%+10) 0.01Ω
2kΩ (0.4%+5) 0.1Ω
20kΩ
200kΩ 10Ω
2MΩ 100Ω
20MΩ (1.2%+25) 1kΩ
200MΩ (5%+45) 10kΩ

Open voltage: less than 3V

Overload protection: DC 250V or AC peak value.

Note:
At range 200Ω, should make the test leads short, and measure the resistance of the wire, then, subtract it from the actual measuring value.

Capacitance

Range Accuracy Resolution
20nF/200nF (4.0%+50) 1pF/10pF
2uF 100pF
20uF 1nF
200uF 10nF
2000uF 100nF

Overload protection: DC 36V or AC peak value

Frequency

Range Accuracy Resolution
20Hz (0.1%+3) 0.001Hz
200Hz 0.01Hz
2kHz 0.1Hz
20kHz 1Hz
200kHz 10Hz
2MHz 100Hz

Input sensitivity: 1.5Vrms.

Overload protection: DC 250V or AC peak value (within 10sec.)

hFE

Range Scope Testing condition
hFE NPN or PNP (0~1000)β Base DC current is approx. 10uA,Vce is approx. 3V

Conductance measurement

Range Accuracy Resolution
(0.1~100)nS (1.0%+30) 0.1nS