AMM-1022 Digital Multimeter

AKTAKOM AMM-1022 Digital Multimeter
Click an image to enlarge
Brand:AKTAKOM
User's Guide:Download

Digital Multimeter Aktakom AMM-1022 is a portable professional measuring instrument with LCD screen foreasy reading. Single operation of a transform switch makes measurement convenient. Overload protection, low battery indication and data hold functions are provided.

Retail price :$19.00

Depending on a destination shipping cost may vary.
Free shipping for orders over $450
on items weighting under 20 lbs
(certain restrictions apply!)

See shipping page for international rates
or contact sales@tmatlantic.com

In stock now

BuyAdd to cart

Digital Multimeter Aktakom AMM-1022 is a portable professional measuring instrument with LCD screen foreasy reading. Single operation of a transform switch makes measurement convenient. Overload protection, low battery indication and data hold functions are provided.

Features:

  • Max. Display: 1999
  • Data Hold
  • Diode Test
  • Continuity Buzzer
  • Transistor hFE test
  • Temperature: -20...1000C

DC Voltage (DCV)1000 V
DCV Accuracy(0.5%+2)
DCV Resolution0.1 mV
AC Voltage (ACV)750 V
ACV Accuracy(1.2%+5)
ACV Resolution1 mV
DC Current (DCA)10 A
DCA Accuracy(1%+2)
DCA Resolution0.1 μA
Resistance200 Ω...20 MΩ
Temperature-20 C...+1000 C
Temperature resolution1 C
Temperature sensor typeType K thermocouple
Diode test+
Continuity check+
Data hold+
DisplayLCD, 1999 counts
Power supply1.5 V x 3 batteries
Dimensions158 x 74 x 31 mm / 6.2 x 2.9 x 1.2 in

Accuracy tables

DC Voltage

RangeResolutionAccuracyAbsolute errorValue-errorValue Learn an absolute error of your measurementValue+error
0.2 V 0.1 mV (0.5%+2 digits) 0.0011995 V V
2 V 1 mV (0.5%+3 digits) 0.012995 V V
20 V 0.01 V (0.8%+3 digits) 0.18992 V V
200 V 0.1 V (0.8%+3 digits) 1.8992 V V
1000 V 1 V (0.8%+5 digits) 12.992 V V

AC Voltage

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
2 V 1 mV Unspecified Unspecified V
20 V 10 mV Unspecified Unspecified V
200 V 0.01 V (1.2%+5 digits) 2.44988 V V
750 V 1 V (1.2%+5 digits) 13.988 V V

DC Current

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 A 0.1 A (1.0%+2 digits) 2.199 A A
2 mA 1 A (1.0%+2 digits) 0.02199 mA mA
20 mA 10 A (1.5%+2 digits) 0.31985 mA mA
200 mA 100 A (1.5%+2 digits) 3.1985 mA mA
10 A 10 mA (3.0%+2 digits) 0.3197 A A

Resistance

RangeResolutionAccuracyAbsolute errorValue-errorValueValue+error
200 Ω 0.1 Ω (1.0%+3 digits) 2.299 Ω Ω
2 kΩ 1 Ω (1.0%+2 digits) 0.02199 kΩ
20 kΩ 10 Ω (1.0%+2 digits) 0.2199 kΩ
200 kΩ 100 Ω (1.0%+2 digits) 2.199 kΩ
2 MΩ 1 kΩ (1.0%+2 digits) 0.02199 MΩ
20 MΩ 10 kΩ Unspecified Unspecified

Specifications:

FunctionRangeResolutionAccuracy
DC Voltage0.2V0.1mV±(0.5%+2)
2V1mV(0.5%+3)
20V0.01V(0.8%+3)
200V0.1V(0.8%+3)
1000V1V(0.8%+5)

Input impedance: 1MΩ
Max. input voltage: 200mV range: 250VDC or AC rms
2V…1000V ranges: 1000VDC or 750V rms

FunctionRangeResolutionAccuracy
AC Voltage2V1mV-
20V10mV-
200V0.01V(1.2%+5)
750V1V(1.2%+5)

Input impedance: 1MΩ
Max. input voltage: 200mV range: 250VDC or AC rms
2V1000V ranges: 1000VDC or 750V rms
Frequency range: 40Hz-400Hz
Response: Average, calibrated in rms of sine wave.

FunctionRangeResolutionAccuracy
DCCurrent200μA0.1μA(1%+2)
2 mA1μA(1%+2)
20mA10μA(1.5%+2)
200mA100μA(1.5%+2)
10A10mA(3.0%+2)

Overload protection: F250mA/250V fuse (10A range unfused).
Max. Input Current: mA Jack: 200mA, 10A Jack: 10A

FunctionRangeResolutionAccuracy
Resistance200Ω0.1Ω(1%+3)
2kΩ(1%+2)
20kΩ10Ω(1%+2)
200kΩ100Ω(1%+2)
2MΩ1kΩ(1%+2)
20MΩ10kΩ-

Overload protection: 250V DC or 250V AC rms.

FunctionRangeResolutionAccuracy
Temperature-20°C0C1C(5.0%+4)
0C400C1C(1.0%+3)
400C1000C1C(2.0%+3)

Overload Protection: 250V DC or rms AC.

Specifications:

  • CATⅢ 600V, CATⅡ1000V
  • Maximum value measurement
  • Data hold
  • Back light
  • Display: 2000 counts, digit is 16mm high
  • Fuselage colour: Gray
  • Weight: Approx. 220g/0.485lb
  • Size: 158 x 74 x 31 mm / 6.2 x 2.9 x 1.2 in
  • Test Leads

  • Thermocouple (K type)

  • Battery: 1.5V AAA x 3

  • Instruction Manual

  • Temperature measurement

    1. Set the transform switch at the TEMP range position.
    2. The LCD display will show the current environment temperature.
    3. When measuring the temperature with thermocouple, ‘K’ type probe for this meter can be used. Insert the black plug to the COM jack and the red one to the TEMP jack, touch the end of the temperature sensor to the area or surface of the object for measurement.
    4. You can get reading from LCD.

    To avoid electrical shock, do not connect the thermocouples with the electriferous circuit.

    Transistor measurement

    1. Set the rotary switch to hFE range.
    2. Determine whether the transistor to be tested is NPN or PNP type and locate the Emitter, Base and Collector leads.
    3. Insert leads of the transistor into proper holes of the hFE socket.
    4. The meter will show the approx. hFE value at test condition of base current 10μA and VCE 2.8V.

    To avoid electrical shock and/or damage to the instrument, before attempting to insert transistors for testing, always be sure that test leads have been disconnected from any measurement circuits.

    Voltage measurement

    1. Set rotary switch to the desired V= or V∼ range.
    2. Connect the black and red test leads to the COM and V terminals respectively.
    3. Connect the test leads to the circuit being measured
    4. Read the displayed value. The polarity of red test lead connection will be indicated when making a DCV measurement.
    5. When only the figure "1" is displayed, it indicates overrange situation and the higher range has to be selected.

    To avoid electrical shock and/or damage to the instrument, do not attempt to take any voltage measurement that might exceeds 1000VDC or 750VAC rms. Do not apply more than 1000VDC or 750VAC rms between the common terminal and the earth ground.

    Continuity test

    1. Set the rotary switch to range.
    2. Connect the black and red test leads to the COM and Ω terminals respectively.
    3. Connect the test leads to the resistance in the circuit being measured.
    4. When the test lead to the circuit is below 50Ω, a continuous beeping will indicate it.

    To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before testing for Continuity.

    Current measurement

    1. Set the rotary switch to the desired A= or A∼ range.
    2. Connect the black test lead to the COM terminal and the red test leads to the mA terminal for a maximum of 200mA. For a maximum of 10A, move the red test lead to the 10A terminal.
    3. Connect test leads in series with the load in which the current is to be measured.
    4. Read the displayed value. The polarity of red test lead connection will be indicated when making a DCA measurement.
    5. When only the figure "1" displayed, it indicates overrange situation and the higher range has to be selected.

    To avoid damage to the meter, use the proper terminals, function, and range for your measurement.

    NOTE:
    - When only the figure ‘OL’ is displayed, it indicates overrange situation and the higher range has to be selected.
    - When the value scale to be measured is unknown beforehand, set the range selector at the highest position.
    - “” means the socket of INPUT maximum current is 200mA, over current will destroy the fuse. 10A´s maximum current is 10A, no fuse protection.

    Resistance measurement

    1. Connect the black test lead to the COM jack and the red test lead to the Ω jack.
    2. Set the transform switch at the Ω range position.
    3. Connect test leads across the resistance under measurement.
    4. You can get reading from LCD.

    To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before measuring resistance.

    NOTE:
    - For measuring resistance above 1MΩ, the meter may take a few seconds to stabilize reading. This is normal for high resistance measuring.
    - If the resistance being measured exceeds the maximum value of the range selected or the input is not connected, an overrange indication "1" will be displayed.

    Diode test

    1. Set the rotary switch to range.
    2. Connect the black and red test leads to the COM and Ω terminals respectively.
    3. Connect the red test lead to the anode, black test lead to the cathode of the diode under testing.
    4. The meter will show the approx. forward voltage of the diode. If the lead connection is reversed, only figure "1" displayed.

    To avoid electrical shock and/or damage to the instrument, disconnect circuit power and discharge all high-voltage capacitors before testing diodes.

    El Multmetro digital Aktakom AMM-1022 es un instrumento porttil de medicin profesional con pantalla LCD fcil de leer. El interruptor de una sola operacion de transformacin hace que la medicin sea cmoda. Proteccin contra sobrecarga, indicador de batera baja y retencin de datos se proporcionan.

    Caractersticas:

    • Max. Pantalla: 1999
    • Retencin de datos
    • Prueba de diodos
    • Buzzer de continuidad
    • Prueba de transistor hFE
    • Temperatura: -20...1000C